A rugged Wet/Wet Differential!
Ultra rugged, intrinsically safe, Differential Pressure Gauge with 0.1% of reading accuracy, capable of continuous vacuum measurement.
From: $3,839.00
A rugged Wet/Wet Differential!
Ultra rugged, intrinsically safe, Differential Pressure Gauge with 0.1% of reading accuracy, capable of continuous vacuum measurement.
The Ametek Crystal XP2i-DP is a utra rugged, intrinsically safe, Differential Pressure Gauge with 0.1% of reading accuracy, capable of continuous vacuum measurement.
| DIFFERENTIAL PRESSURE GAUGE ACCURACY | |
|---|---|
| 0 to 20% of Range: ±(0.02% of Full Scale + Static Line Effect*) | Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. |
| 20% to 110% of Range: ±(0.1% of Reading + Static Line Effect*) | For continuous vacuum use, connect to the right (-) port marked “Vacuum this side.” |
| *Static Line Effect: 0.0003 psi/psi lowest static pressure. | |
| OPERATING TEMPERATURE | |
| Temperature Range: -10 to 50° C (14 to 122° F) | Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification. |
| DISPLAY | |
| Screen: 5.5 digits, 16.9 mm (0.67″) high | LCD readable in sunlight with bright backlight. |
| Display Rate: 4 readings/second (standard) 8 readings/second (PSV mode) |
PSV mode is intended for Pressure Safety Valve testing, and requires software to enable. (ConfigXP is a free downloadfrom ametekcalibration.com.) |
| POWER | |
| 3 x AA: 1500 hours, typical | Uses 3 alkaline AA (LR6) batteries. Use of backlight and/or RS232 reduces operating time. |
| Ultra Low Power: > 1 year, typical** | Optional AC power supply available. |
| For hazardous location product warnings, refer to the operation manual. |
|
| **20 minute recording interval or greater in LT5 mode. | |
| DATA/COMMUNICATION | |
| Digital Interface: RS232, DB-9F | Free Labview® drivers and programming instructions are available for download from website. |
| For hazardous location product warnings, refer to the operation manual. |
|
| DATALOGGING | |
| Capacity: 32 000 data points | DataLoggerXP Firmware Update may be purchased at time of order, or an Upgrade Package may be purchased afterward. |
| Storage Type: Non-volatile flash memory | Software available for download from Ametek/Crystal website. Compatible with 32 & 64 bit Windows 7, Vista, and XP (32 bit only). Produces csv files, or uses Excel template files (samples included) to automatically format and graph data. |
| Minimum Interval: 1 second | |
| Maximum Interval: 18 hours | |
| Min. Event Duration: 250 mS*** | *** Logging Type – Average with Peaks. |
| psi Versions | 15PSIXP2I-DP 15 psi Differential Pressure / Vacuum Gauge, 100PSIXP2I-DP 100 psi Differential Pressure / Vacuum Gauge |
|---|---|
| DataLogging Option | None, -DL DataLoggerXP Firmware Upgrade. Includes 3313 USB adapter cable |
| Fitting Location | Bottom (standard), -RP Rear Pressure Fitting, -F4 Panel Mount Flange – Includes the -RP option |
| Other Options |